CT CHEST/ABDOMEN/PELVIS C+
(last updated 6/20/2016, last reviewed 02/07/2022)
(last updated 6/20/2016, last reviewed 02/07/2022)
INDICATION: Metastases, lymphoma, adenopathy, mass, unexplained weight loss.
SERIES
- 3mm Axial Chest Soft Tissue
- 1mm Axial Lung
- 3mm Axial A/P
- 3mm Coronal chest only (Lung windows)
- 3mm Sagittal chest only (Lung windows)
- 3mm Coronal A/P only
- 3mm Sagittal A/P only
| CHEST | ABD/PEL |
POSITIONING/MODE
|
supine/helical |
supine/helical |
kV/mA/rot time
|
120kV/auto mA/0.5 sec |
120kV/auto mA/0.5 sec |
DETECTOR COLLIMATION
|
16 x 1.5mm (S16) 24 x 1.2mm (D64) |
16 x 1.5mm (S16) 24 x 1.2mm (D64) |
SLICE THICKNESS
|
3mm / 1mm |
3mm |
PITCH
|
1 to 1.5 |
1 to 1.5 |
TABLE FEED
|
4.8 cm/sec (S16) 5.8 cm/sec (D64) |
4.8 cm/sec (S16) 5.8 cm/sec (D64) |
KERNAL
|
std and lung |
std |
SFOV
|
large |
large |
COVERAGE
|
thoracic inlet -> diaphragm |
diaphragm -> pubic symphysis |
ACQUISITION
|
cradniocaudal |
cradniocaudal |
ORAL CONTRAST
|
yes |
yes |
IV CONTRAST
|
100 mL Isovue 300 30 mL saline |
yes |
INJECTION RATE
|
2.5 mL/sec |
--- |
SCAN DELAY
|
Care bolus at ascending aorta (threshold 100 HU), ~ 15 sec |
30 sec delay b/w chest and A/P (S16) 33 sec delay b/w chest and A/P (D64) |
RESPIRATION
|
Inspiration |
Inspiration |
POSTPROCESSING
|
3mm Cor and Sag (lung windows, chest only) |
3mm Cor and Sag (A/P only) |
NOTES:
***For pediatric patients (<18 yo) follow ALARA principle and use pediatric specific technical parameters per respective CT scanner***